Demo Instructions
6
Warning: defibrillators/external pacemakers are capable of emitting lethal
high voltages/energies and should be operated only by trained professionals.
3. The analyzer calculates the energy delivered over the load resistance and displays the
result in joules under the result. Press F4 for summary
4. Show on the monitor that ECG waveform is converted from V. Fib. to Normal Sinus
Rhythm.
.
Ventricular Fibrillation
Normal Sinus Rhythm
5. Following the same procedures if testing an AED (Automated External Defibrillator