Impulse 6000D
Defibrillator Analyzer
Impulse 7000DP
Defibrillator/Transcutaneous Pacer Analyzer
Users Manual
August 2007, Rev. 1, 1/08
© 2007 Fluke Corporation, All rights reserved.
All product names are trademarks of their respective companies.

Table of Contents

Other manuals for Fluke Biomedical IMPULSE 6000D

Question and Answer IconNeed help?

Do you have a question about the Fluke Biomedical IMPULSE 6000D and is the answer not in the manual?

Fluke Biomedical IMPULSE 6000D Specifications

General IconGeneral
BrandFluke Biomedical
ModelIMPULSE 6000D
CategoryMeasuring Instruments
LanguageEnglish

Summary

Warranty and Product Support

Warranty Details

Outlines the terms and conditions of the product warranty.

Notices

All Rights Reserved

States copyright and usage restrictions for the publication.

Copyright Release

Grants limited permission for reproducing manual content for training.

Unpacking and Inspection

Provides instructions for receiving and inspecting the instrument upon arrival.

Technical Support and Claims

Offers contact information for technical assistance and procedures for filing claims.

Standard Terms and Conditions

Refunds and Credits Policy

Details eligibility and conditions for product returns, refunds, and credits.

Restocking Charges

Specifies the fees applied for product returns based on the return timeframe.

Return Procedure

Outlines the steps required for returning instruments for service or credit.

Restrictions and Liabilities

Document Information and Liability

States that information is subject to change and limits Fluke Biomedical's responsibility.

Manufacturing Location

Identifies the place of manufacture for the Impulse 6000D and 7000DP Analyzers.

Safety Information

Warning and Caution Definitions

Explains the meaning of Warning and Caution symbols and their implications.

General Safety Guidelines

Provides essential safety precautions for operating the Analyzer to prevent injury.

Specific Operational Safety Precautions

Details critical safety measures related to power, patient connection, and voltage handling.

Instrument Familiarization

Top-Panel Controls and Connections

Identifies and describes the controls and connectors on the top surface of the Analyzer.

Rear-Panel Connections

Details the input/output ports and connectors located on the rear panel of the Analyzer.

Turning the Analyzer On and Off

Explains the procedure for powering the Analyzer on and off, including initial charging.

Analyzing Defibrillators

Testing Energy Levels

Measures the energy output accuracy of defibrillators.

Changing Waveform Characteristics

Allows modification of test waveform parameters like amplitude and rate.

Testing Defibrillator Synchronization

Verifies the defibrillator's synchronization response to ECG signals.

Testing Defibrillator Charge Time

Measures the time taken for a defibrillator to reach a specified charge level.

Analyzing Pacemakers (7000DP only)

Setting Up for Pacer Testing

Guides the user through configuring the Analyzer for pacemaker analysis.

Performing Pacer Asynchronous Test

Verifies continuous mode pacemaker output with simulated ECG signals.

Performing Pacer Demand Test

Checks demand mode pacemaker interaction with simulated ECG signals.

Performing Pacer Sensitivity Test

Determines the ECG signal amplitude required to trigger the pacemaker.

Performing Pacer Refractory Period Test

Assesses pacemaker's ability to sense ECG activity after paced or sensed events.

Simulating ECG Signals

Connecting to ECG Terminals

Explains how to connect ECG leads from monitors to the Analyzer's posts.

Setting Normal Sinus Rhythm ECG Signal

Configures a standard ECG signal with adjustable rate and amplitude.

Setting Performance ECG Signal

Generates special test signals for defibrillator monitor performance evaluation.

Setting Interactive ECG/Pacer Demand Mode Simulation

Simulates varying ECG conditions for testing pacer response in demand mode.

Setting Interactive ECG/Pacer Asystole Mode Simulation

Simulates asystole conditions for testing pacer response to simulated heartbeats.

Selecting ECG Arrhythmias

Allows simulation of various cardiac arrhythmia waveforms for testing.

Selecting TV Paced Waveforms

Configures pacing signals for testing transvenous pacing modes.

Testing R Wave Detection

R Wave Detection Parameters

Details how the Analyzer simulates R-Waves and allows adjustable parameters.

Noise Immunity Test

Performing Noise Immunity Test

Verifies the pacemaker's ability to filter AC line frequency noise.

Analyzer Setup Functions

Setting Up the Battery

Configures battery options including power off, charging, and training.

Setting Display and Sound Options

Adjusts display contrast, backlight, and beeper volume settings.

Displaying Instrument Information

Shows manufacturing date, firmware version, and serial number.

Controlling the Analyzer Remotely

Using Ansur Test Automation

Describes how to control the Analyzer remotely using Ansur software.

Maintenance

Cleaning the Analyzer

Provides instructions and precautions for cleaning the Analyzer's exterior.

Maintaining Peak Battery Condition

Offers guidance on how to maintain optimal battery capacity over time.

Accessories

Analyzer Accessories List

Lists available accessories for the Impulse 6000D/7000DP Analyzer.

Specifications

General Specifications

Lists general operating and physical characteristics of the Analyzer.

Defibrillator Analyzer Specifications

Details the technical specifications for defibrillator testing functions.

Synchronization Test Specifications

Provides specific parameters and measurements for synchronization testing.

ECG Waveform Specifications

Outlines the parameters and capabilities for simulating ECG waveforms.

Transcutaneous Pacemaker Analyzer Specifications

Details the technical specifications for the 7000DP's pacemaker testing.

Pacer Test Specifications

Lists specifications for various pacemaker tests like pulse rate, width, and energy.

Sensitivity and Refractory Period Tests Specs

Provides specifications for pacer sensitivity and refractory period measurements.

Appendix A: Impulse 6000D/7000DP Remote Operation

Ansur Test Guide

Explains how to use the Ansur plug-in software for executing tests on the Analyzer.

Defibrillator Tests via Ansur

Details how to run defibrillator tests using the Ansur software interface.

Pacemaker Tests via Ansur

Describes how to perform pacemaker tests using the Ansur software.

ECG Waveform Simulation Tests via Ansur

Explains how to simulate ECG waveforms for testing using Ansur.

Battery Performance Tests via Ansur

Details how to conduct battery performance tests using the Ansur software.

Appendix B: Impulse 6000D/7000DP Test Templates

Creating Test Templates

Guides the user on creating, modifying, and reviewing test templates in Ansur.

Using Defibrillator Test Elements

Energy Measurement Test Elements

Describes parameters and data for the Energy Measurement test.

Charge Time Test Elements

Details parameters and measurements for the Charge Time test.

Synchronization Time Test Elements

Outlines parameters and measurements for the Synchronization Time test.

Using Pacemaker Test Elements (Impulse 7000DP only)

Pacer Parameter Test Elements

Explains parameters and measurements for the Pacer Parameter test.

Pacer Refractory Test Elements

Details parameters and measurements for Pacer Refractory Period tests.

Pacer Sensitivity Test Elements

Outlines parameters and measurements for Pacer Sensitivity tests.

ECG Pacer Interactive Test Elements

Describes parameters for simulating patient response to pacemakers.

Pacer Demand Mode Test Elements

Details parameters for testing demand mode pacing over a range of BPM rates.

Related product manuals