EasyManua.ls Logo

Fluke Biomedical IMPULSE 6000D - Defibrillator Analyzer Specifications

Fluke Biomedical IMPULSE 6000D
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Defibrillator/Transcutaneous Pacemaker Analyzer
Specifications
31
Defibrillator Analyzer Specifications
Energy Output Measurement
Compatible Defibrillator Waveshapes.................Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic
Note
AC Pulsed Bi-Phasic waveform has not been approved in the United States.
Autoranged Measurement....................................0.1 to 600 J
Accuracy
0.1 to 360 J .....................................................±(1 % of reading + 0.1 J)
360 to 600 J ....................................................±(1 % of reading + 0.1 J), typical
Note
For Pulsed Bi-Phasic defibrillator, specified accuracy is
±
(1.5 % of reading + 0.3 J) on both ranges.
Load resistance
Resistance...........................................................50 Ω
Accuracy..........................................................±1 %, non-inductive (<2 μH)
Pulse trigger level .................................................20 V
Pulse width
Range..................................................................1.0 to 50.0 ms
Accuracy..............................................................±0.1 ms
Voltage
Range..................................................................20 to 5000 V
Accuracy..............................................................±(1 % of reading + 2 V)
Current
Range..................................................................0.4 to 100.0 A
Accuracy..............................................................±(1 % of reading + 0.1 A)

Table of Contents

Other manuals for Fluke Biomedical IMPULSE 6000D

Related product manuals