Impulse 6000D, 7000DP
Users Manual
40
Sensitivity Test
Automatic Interactive Threshold Detection
Compatible pacer rates.......................................30 to 120 PPM
ECG R wave:
Waveforms..........................................................Square, Triangle, Sine
Width...................................................................1 to 19 (by 1) ms
20 to 95 (by 5) ms
100 to 300 (by 25) ms
Accuracy .............................................................± 5% of setting
Amplitude ............................................................0.05 to 0.95 (by 0.05) mV
1.0 to 5.0 (by 0.5) mV
Accuracy .............................................................± 5% of setting
Refractory Period Tests
Paced Refractory Period ......................................20 to 500 ms
Sensed Refractory Period ....................................15 to 500 ms
Accuracy................................................................±1 ms
Pacer pulse rate ....................................................20 to 200 PPM
ECG
Waveform............................................................Triangle wave
Pulse width..........................................................40 ms
Amplitude ............................................................1.0 mV

Table of Contents

Other manuals for Fluke Biomedical IMPULSE 6000D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke Biomedical IMPULSE 6000D and is the answer not in the manual?

Fluke Biomedical IMPULSE 6000D Specifications

General IconGeneral
BrandFluke Biomedical
ModelIMPULSE 6000D
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals