40
Sensitivity Test
Automatic Interactive Threshold Detection
Compatible pacer rates.......................................30 to 120 PPM
ECG R wave:
Waveforms..........................................................Square, Triangle, Sine
Width...................................................................1 to 19 (by 1) ms
20 to 95 (by 5) ms
100 to 300 (by 25) ms
Accuracy .............................................................± 5% of setting
Amplitude ............................................................0.05 to 0.95 (by 0.05) mV
1.0 to 5.0 (by 0.5) mV
Accuracy .............................................................± 5% of setting
Refractory Period Tests
Paced Refractory Period ......................................20 to 500 ms
Sensed Refractory Period ....................................15 to 500 ms
Accuracy................................................................±1 ms
Pacer pulse rate ....................................................20 to 200 PPM
ECG
Waveform............................................................Triangle wave
Pulse width..........................................................40 ms
Amplitude ............................................................1.0 mV