Fluke Biomedical IMPULSE 6000D - Pacer Test Specifications

Fluke Biomedical IMPULSE 6000D
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Defibrillator/Transcutaneous Pacemaker Analyzer
Specifications
39
Pulse Rate
Range..................................................................5.0 to 800 PPM
Accuracy..............................................................±(0.5% of reading + 0.1 PPM)
Pulse Width
Range..................................................................1.00 to 100.0 ms
Accuracy..............................................................±(0.5% of reading + 0.01 ms)
Energy
Range..................................................................1 µJ to 2.00 J
Accuracy..............................................................±(4% of reading + 10 µJ)
Demand and Asynchronous Mode Test
Input Pacer pulse rates.........................................30 to 200 PPM
ECG NSR wave
Rate.....................................................................10 to 300 (by 1) BPM
Amplitude ............................................................1 mV
Underdrive rate....................................................10 BPM minimum
Overdrive rate......................................................300 BPM maximum