Demo Instructions
7
Warning: defibrillators/external pacemakers are capable of emitting lethal
high voltages/energies and should be operated only by trained professionals.
THE SYNCHRONIZATION TEST (Cardioversion)
Measures the time delay in milliseconds (ms) between the peak of the R wave and the
discharging of the defibrillator pulse. A wide range of simulated ECG waveforms are
available in the analyzer for performing this test. Atrial fibrillation (AFIB) is the most
representative clinical waveform for demonstration purposes. Two selections are
available; AFIB-Coarse is the default simulation.
1. Press F2“Sync”
2. Press F1 “Set Wave” Use UP/DOWN arrow keys to select an ECG waveform. AFIB
is chosen for the purpose of this demonstration.
3. Note ability to change AMPLITUDE and RATE. Select F5 for DONE.
4. Select F2 to select WAVE CONVERSION. Three choices are listed: CONVERT, NO
CONVERT and ASYSTOLE. Convert is chosen for the purpose of this demonstration

Other manuals for Fluke Biomedical IMPULSE 6000D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke Biomedical IMPULSE 6000D and is the answer not in the manual?

Fluke Biomedical IMPULSE 6000D Specifications

General IconGeneral
BrandFluke Biomedical
ModelIMPULSE 6000D
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals