EasyManua.ls Logo

Fluke 323 - Page 14

Fluke 323
43 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
323/324/325
Calibration Manual
14
Table 4. Performance Tests (cont.)
Test
(Switch
Position)
Calibrator
Output
323 324 325
Meter Reading Limit
Low High
AC Amps
(with 50-
Turn Coil)
0.12 A @ 50 Hz X N/A N/A 5.4 A 6.6 A
0.48 A @ 150 Hz X N/A N/A 22.9 A 25.1A
4.2 A @ 50Hz X N/A N/A 205.3 A 214.7 A
7.8 A @ 50 Hz X N/A N/A 381.7 A 398.3 A
7.8 A @ 150 Hz X N/A N/A 379.7 A 400.3 A
7.8 A @ 400 Hz X N/A N/A 379.7 A 400.3 A
0.004 A @ 50 Hz N/A X N/A 0.15 A 0.25 A
0.2 A @ 50 Hz N/A X N/A 9.80 A 10.20 A
0.6 A @ 50 Hz N/A X N/A 29.50 A 30.50 A
2 A @ 50 Hz N/A X N/A 97.5 A 102.5 A
6 A @ 50 Hz N/A X N/A 293.5 A 306.5 A
0.2 A @ 50 Hz N/A N/A X 9.75 A 10.25 A
0.6 A @ 50 Hz N/A N/A X 29.35 A 30.65 A
2 A @ 50 Hz N/A N/A X 98.0 A 102.0 A
6 A @ 50 Hz N/A N/A X 295.0A 305.0 A
DC Amps
(with 50-
Turn Coil)
0.2 A N/A N/A X 9.75 A 10.25 A
0.6 A N/A N/A X 29.35 A 30.65 A
2 A N/A N/A X 97.5 A 102.5 A
6 A N/A N/A X 293.5 A 306.5 A

Other manuals for Fluke 323

Related product manuals