EasyManua.ls Logo

Fluke 433 - Page 34

Fluke 433
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 433, 434, 435
Service Manual
3-10
Accuracy Check of Channel N (Neutral):
11. Connect the N input with the C/L3 input (See Figure 3-4). A/L1, B/L2, C/L3 now
give zero reading.
12. Set the Calibrator to 0 Hz / 0 V and then to OPR.
13. Fluke 433/434: check for a voltage readout V rms between 0.0 ... 0.6 V.
Fluke 435: check for a voltage readout V rms between 0.00 ... 0.12 V.
14. Check the 120 V range according to the table below.
Table 3-4. Accuracy Check of Voltage Channel N (Neutral)
Set Calibrator to Readout at Voltage Channel N
0 Hz, 0 V, OPR Fluke 433/434: 0.0 ... 0.6 V, Fluke 435: 0.00 ... 0.12 V
60 V, 60 Hz, OPR Fluke 433/434: 59.4 ... 60.6 V, Fluke 435: 59.88 ... 60.12 V
120 V, 60 Hz, OPR Fluke 433/434: 119.4 ... 120.6 V, 59.99 ... 60.01 Hz
Fluke 435: 119.88 ... 120.12 V, 59.990 ... 60.010 Hz
240 V, 60 Hz, OPR Fluke 433/434: 239.4 ... 240.6 V, 59.99 ... 60.01 Hz
Fluke 435: 239.88 ... 240.12 V, 59.990 ... 60.010 Hz
15. Set the Calibrator to STBY.
Optional test. Bandwidth Check of Channel N (*):
16. Set the Calibrator to 120 V, 60 Hz and OPR.
17. Now check the voltage channel N. Use the Calibrator’s field edit function (5500A:
FIELD EDIT key, 5700A: AMPL/FREQ key) to adjust the Calibrator to an Analyzer
readout of 120.0 V / 120.00 V.
18. Increase the frequency to 3 kHz.
Fluke 433/434: check for a readout of 113.8 V or more.
Fluke 435: check for a readout of 114.00 V or more.
19. Set the Calibrator to STBY.
3.7.3 Verification of voltage inputs in 230 V range
Accuracy Check of Channel A/L1, B/L2, C/L3:
1. Set the Analyzer to 230 V, 50 Hz nominal (SETUP menu, arrow and ENTER keys).
Press MENU, select Volts/Amps/Hertz, press F5 - OK.
2. To check the A/L1, B/L2, C/L3 inputs: connect the N input to Ground (See Figure 3-
3).
3. Set the Calibrator to 0 Hz, 0 V and then to OPR.
4. Fluke 433/434: check for a voltage readout V rms between 0.0 ... 1.2 V.
Fluke 435: check for a voltage readout V rms between 0.00 ... 0.23 V.
5. Check the 230 V range according to the table below.

Related product manuals