EasyManua.ls Logo

Fluke 433 - Page 52

Fluke 433
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 433, 434, 435
Service Manual
4-10
2. The display must show step CL 0300.
If it does not, then press
F1
or
F2
to select the first calibration step in Table 4-
1.
3. Set the Calibrator output to source the signal 0 Hz, 0 V required for the calibration
steps in Table 4-1.
4. Set the Calibrator in operate (OPR) or standby (STBY) as indicated.
5. Press
F3
to start the calibration.
6. Wait until the display shows calibration status
:READY .
7. Press
F2
to select the next calibration step, and start the calibration.
Continue through all calibration points of Table 4-1.
8. When you are finished, set the Calibrator to Standby (STBY).
9. Continue at Section 4.6.2.
Table 4-1. Offset Adjustment of all Inputs
Cal step Description Calibrator Setting
CL 0300 OffsetLowVolt 0 V, 0 Hz, OPR
CL 0310 Offset125Volt 0 V, 0 Hz, OPR
CL 0320 Offset250Volt 0 V, 0 Hz, OPR
CL 0330 Offset500Volt 0 V, 0 Hz, OPR
CL 0340 Offset6KVolt 0 V, 0 Hz, OPR
--- --- STBY
4.6.2 Low voltage and current gain adjustment
Proceed as follows to do the Low Voltage and Current Gain Adjustment:
1. Press
F2
to select calibration step CL 0400.
2. Keep the Analyzer connected to the Calibrator as shown in Figure 4-3.
3. Set the Calibrator to source 0.67 V, 50 Hz .
4. Set the Calibrator to operate (OPR).
5. Press
F3
to start the calibration.
6. Wait until the display shows calibration status
:READY .
7. Set the Calibrator to Standby (STBY).
4.6.3 Voltage gain adjustment
Proceed as follows to do the Voltage gain Adjustment.
1. Press
F2
to select the first calibration step in Table 4-2 (CL 0410).
2. Connect the Analyzer to the Calibrator as shown in Figure 4-4: disconnect the current
inputs!

Related product manuals