71X Series
Calibration Manual
12
Firmware V2.0 and Later
Table 13. 714 Temperature Measure and Thermocouple Simulate
TC Type Range °C Accuracy °C *
J
-210.0 to 0.0 0.6
0.0 to 800.0 0.4
800.0 to 1200.0 0.5
K
-200.0 to 0.0 0.8
0.0 to 1000.0 0.5
1000.0 to 1372.0 0.7
T
-250.0 to 0.0 0.8
0.0 to 400.0 0.4
E
-250.0 to -100.0 0.8
-100.0 to 1000.0 0.4
R
-20.0 to 0.0 2.0
0.0 to 1787.0 1.4
S
-20.0 to 0.0 2.0
0.0 to 1767.0 1.4
B
600.0 to 800.0 1.4
800.0 to 1000.0 1.5
1000.0 to 1820.0 1.7
L
-200.0 to 0.0 0.45
0.0 to 900.0 0.4
U
-200.0 to 0.0 0.7
0.0 to 600.0 0.45
Maximum input voltage: 30 V
* Includes 0.2 °C cold junction compensation (CJC) error
Table 14. 714 Millivolt Measure and Source
Range Resolution Accuracy
-10 mV to 75 mV 0.001 mV 0.015 % ± 10 µV
Maximum input voltage: 30 V
Maximum source current is 1.0 mA