718 Series
Product Overview
4
XW Warning
To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery
as soon as the battery indicator B appears.
To avoid a violent release of pressure in a pressurized system, shut off the valve and slowly bleed off the
pressure before you attach or detach the internal pressure sensor or Pressure Module fitting to the
pressure line.
To avoid over pressure damages, do not apply pressure that exceeds the limits listed in the Pressure
Specifications table in the “Specifications” section.
To avoid mechanically damaging the Calibrator, do not apply torque between the pressure fitting and the
Calibrator case. See Figure 1 for the proper use of tools.
To avoid misleading readings, disconnect the Pressure Module connector at the Calibrator.
To avoid damage to the Pressure Module, refer to the related Instruction Sheet.
To avoid damage to the pump, use with dry air and non-corrosive gases only.
Check test leads for continuity before using. Inspect Calibrator for cracks or damage, do not use the probes
if they are damaged or show high resistance.