Troubleshooting
Logger Communications Problems
Communications lead not fully inserted – Check correct sockets are
being used.
Damaged communications lead or connectors – Check for breaks
and other damage; replace the lead.
Batteries not charged – Recharge the batteries, ensuring the charging
LEDs are illuminated (see p. 21).
Try a hardware reset using the default reset parameters (see p. 26).
Logger Download Error Messages
Error Message Action
There are
insufficient readings
in the logger
Check trigger set point (time or temperature).
Check logger’s battery for charge.
Check date/time settings on computer.
Check probes and their connections.
Reset logger and test probes (see ‘Logger Diagnostics’, below).
Logger stopped due
to going over
temperature
The logger’s maximum-permitted internal temperature has been
exceeded and it may have suffered serious damage: contact Datapaq
for advice.
Logger stopped due
to low battery
Replace or recharge the battery as appropriate, then repeat the
profile run.
Logger memory full Data collection may have stopped before the run was completed:
check the data collection period and sample interval before resetting
the logger for another run (see ‘Resetting the Data Logger’, p. 46).
Checking the Data
Thermocouple probes are generally reliable, but damage resulting from
inappropriate use or handling can produce erroneous readings. If you suspect
that invalid data may have been introduced into your temperature profile
(paqfile), select the View Data tab in the Insight software’s Analysis Window to
view the raw data as downloaded from the logger. The various types of invalid
data which may be contained in a paqfile are shown in the analysis grid as
follows.
*OC* Open circuit.
*NA* Telemetry data not received.