PV350 English Instruction Sheet
Page 1
®
PV350
Pressure/ Vacuum Transducer Module
PV350
PRESSURE / VACUUM MODULE
cmHg
inHg
kPa
psi
OFF
M
ET
R
IC
E
N
G
L
IS
H
Z
E
Instruction Sheet
Introduction
The PV350 Pressure/Vacuum Transducer Module (the module)
measures gauge pressure; the difference between atmospheric
(barometric) pressure and the pressure (or vacuum) applied to the
pressure port on the transducer. It then converts that measurement to 1
mV dc per unit. A module and a high-performance digital multimeter
(DMM) with min/max function become a datalogger capable of measuring
peak high and peak low.
Box Contents
Pressure/Vacuum Transducer Module
1/4 in. male flare to 1/4 in. female pipe thread connector
9 V battery
Instruction Sheet
Quick Reference Guide
Safety Information
WWarning
To avoid eye injury, always wear approved safety glasses
when working with this instrument.
To avoid injury from pressure explosion:
Be sure that pressure is removed before
disconnecting any hoses or tubing from the PV350
transducer.
Follow accepted safety procedures for systems on
which pressure or vacuum measurements are made.
Make sure the PV350 is serviced by a qualified
technician.
Complies with safety standard EN/IEC 61010.
PN 926873 May 1992 Rev. 5, 1/07
© 1999-2007 Fluke Corporation, All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies.

Other manuals for Fluke PV350

Question and Answer IconNeed help?

Do you have a question about the Fluke PV350 and is the answer not in the manual?

Summary

Introduction to PV350 Module

Safety Information

Materials Compatibility

Cleaning Procedures

Measuring Pressure and Vacuum

Testing and Replacing Battery

Testing the Battery

Procedure to check the battery voltage using the DMM and PV350 module.

How to Replace the Battery

Step-by-step instructions for safely replacing the 9V battery in the module.

Typical Applications

Conversion Factors and Calibration

Conversion Factors

Table of conversion factors for pressure units like psi, kPa, cm-Hg, and inches-Hg.

How to Calibrate the PV350

Detailed steps for calibrating the PV350 module annually for performance.

How to Verify Calibration

Specifications

Pressure and Vacuum Specifications

Details accuracy and range for pressure (up to 500 psi) and vacuum (up to 29.9 in-Hg).

Output and Resolution

Specifies the module's output signal (1 mV/unit) and measurement resolution.

RFI Specification and Impedance

Describes RFI susceptibility and meter input impedance requirements.

Operating Environment and General Information

Operating Environment

Defines acceptable temperature and humidity ranges for module operation.

Temperature Derating

Explains accuracy derating factors based on ambient temperature.

General Specifications

Lists module weight, battery type, battery life, vibration, shock, and cable details.

Summary

Introduction to PV350 Module

Safety Information

Materials Compatibility

Cleaning Procedures

Measuring Pressure and Vacuum

Testing and Replacing Battery

Testing the Battery

Procedure to check the battery voltage using the DMM and PV350 module.

How to Replace the Battery

Step-by-step instructions for safely replacing the 9V battery in the module.

Typical Applications

Conversion Factors and Calibration

Conversion Factors

Table of conversion factors for pressure units like psi, kPa, cm-Hg, and inches-Hg.

How to Calibrate the PV350

Detailed steps for calibrating the PV350 module annually for performance.

How to Verify Calibration

Specifications

Pressure and Vacuum Specifications

Details accuracy and range for pressure (up to 500 psi) and vacuum (up to 29.9 in-Hg).

Output and Resolution

Specifies the module's output signal (1 mV/unit) and measurement resolution.

RFI Specification and Impedance

Describes RFI susceptibility and meter input impedance requirements.

Operating Environment and General Information

Operating Environment

Defines acceptable temperature and humidity ranges for module operation.

Temperature Derating

Explains accuracy derating factors based on ambient temperature.

General Specifications

Lists module weight, battery type, battery life, vibration, shock, and cable details.

Fluke PV350 Specifications

General IconGeneral
CategoryTransducer
ModelPV350
ManufacturerFluke
Measurement TypePressure
Accuracy±0.05 % of full scale
Media CompatibilityCompatible with most non-corrosive gases and liquids
Warranty1 year
Connection1/4 in NPT