SUPER-DAQ Precision Temperature Scanner
Required Equipment
5
Required Equipment
Table 2 lists the equipment required for performance tests and calibration of the
Product.
Table 2. Required Test Equipment
Function Instrument Type Model Comments
Volts DC
Standard Fluke 5720A
8½ digit meter Fluke 8508A
Used to characterize the
5522A
4-wire short
Fluke 884X-SHORT, low
thermal 4-wire short or
equivalent
Fluke PN 2653346
Alternate standard
[1]
Fluke 5522A
Must be characterized with
8508A
Ohms
Standard Fluke 5720A
4-wire short
Fluke 884X-SHORT, low
thermal 4-wire short or
equivalent
Fluke PN 2653346
Alternate standard
The standard resistors
which have equivalent
spec with Fluke 5720A.
PRT
Standard
Fluke 5522A (for 1 kΩ
value)
Standard resistors of
25 Ω, 100 Ω, 200 Ω and
400 Ω
Resistance accuracy
specification (k = 2) should
be equal to or better than:
6 ppm (for 25 Ω)
4 ppm (for 100 Ω, 200 Ω
and 400 Ω)
4-wire short
Fluke 884X-SHORT, low
thermal 4-wire short or
equivalent
Fluke PN 2653346
Alternate standard
The standard resistors
which have equivalent
specification
Thermistor
Standard Fluke 5522A
Must be characterized with
8508A
8½ digit meter Fluke 8508A
Used to characterize the
5522A
4-wire short
Fluke 884X-SHORT, low
thermal 4-wire short or
equivalent
Fluke PN 2653346
Alternate standard
[1]
Fluke 5720A