SUPER-DAQ Precision Temperature Scanner
Calibration Adjustment
27
Command References
CALibrate:ABORt
Description: Instruct unit to abort calibration procedure after present
step.
Example: CAL:ABOR
Related Commands:
CALibrate:STATe?
CALibrate:BACKup
Description: Backup to previous entry point in calibration procedure.
Example: CAL:BACK
Related Commands:
CALibrate:SKIP
CALibrate:SECTion
CALibrate:CONSt? “<cco_name>”
Description: Retrieves the value in use of the given calibration constant.
Example: CAL:CONS? “DC100MV_A1”
Response: -1.401686110719e-04
Related Commands:
CALibrate:SECure:STATe
CALibrate:STORe
CALibrate:DATE?
Description: Return a CAL date associated with stored calibration
constants.
Example: CAL:DATE?
Response: 2013,11,1
Related Commands:
CALibrate:MODule:DATE?