1586A
Calibration Manual
28
CALibrate:INFOrmation?
Description: Return message or instructions associated with
running step
Example: CAL:INFO?
Response: “Connect calibrator to Volt terminals”
Related Commands:
CALibrate:STATe?
CALibrate:STEP?
CALibrate:MODule:DATE <slot>,<year>,<month>,<day>
CALibrate:MODule:DATE? <slot>
Description: Set and query module calibration date.
Example: CAL:MOD:DATE 1,2013,11,1
CAL:MOD:DATE? 1
Response: 2013,11,1
Related Commands:
CALibrate:SECure:STATe
CALibrate:NEXT [<reference>]
Description: Continue a calibration procedure if it is waiting. An
optional parameter reference value (used if it's waiting
for a reference), If the reference value has no unit, the
unit is assumed to be that returned by the CAL_REF?
command.
Example: CAL_NEXT
CAL_NEXT 2.999987
CAL_NEXT 100 mV
Related Commands:
CALibrate:REFerence?
CALibrate:BACKup
CALibrate:SKIP