SUPER-DAQ Precision Temperature Scanner
Calibration Adjustment
29
CALibrate:REFerence?
Description: Return nominal value expected for reference entry
Example: CAL:REF?
Response: 3.000000e+00
Related Commands:
CALibrate:NEXT
CALibrate:STATe?
CALibrate:INFOrmation?
CALibrate:STEP?
CALibrate:SECure:STATe <boolean>, <admin_password>
Description: Instruct Product to enable calibration
Example: CAL:SEC:STATE OFF,”1586”
Related Commands:
CALibrate:SECure:STATe?
CALibrate:SECure:STATe?
Description: Query calibration enable state
Example: CAL:SEC:STAT?
Response: 1
Related Commands:
CALibrate:SECure:STATe
CALibrate:SECTion
Description: Skip to next section of calibration procedure.
Example: CAL:SECT
Related Commands:
CALibrate:BACKup
CALibrate:SKIP
CALibrate:SKIP
Description: Skip to next entry point in calibration procedure.
Example: CAL:SKIP
Related Commands:
CALibrate:BACKup
CALibrate:SECTion