FONA XPan 3D Operating Instructions
26/28 69 683 70110 - 150831
13. Electromagnetic Compatibility
13.1 Electromagnetic emission
The FONA XPan 3D system is suitable for use in the electromagnetic environment specified below. The
customer or the user of the FONA XPan 3D system should assure that it is used in such an environment.
Electromagnetic environment
Radiated and conducted RF
emissions
The FONA XPan 3D system uses RF energy only for its
internal functions. Therefore, the RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
The FONA XPan 3D is suitable for use in domestic
environment and in environments directly connected to
the low voltage power supply network which supplies
buildings used for domestic purposes.
Harmonic Emissions
EN 61000-3-2
Voltage fluctuations/ flicker
emissions 61000-3-3
13.2 Electromagnetic immunity
The FONA XPan 3D system is suitable for use in the electromagnetic environment specified below. The
customer or the user of the FONA XPan 3D system should assure that it is used in such an environment.
Electromagnetic
environment
Electrostatic discharge (ESD)
EN 6 1000-4-2
Non-life-supporting equipment
3 V/m
80 Mhz to 2.5 GHz
Life-supporting equipment
10 V/m
80 MHz to 2.5 GHz
Conducted RF
EN 61000-4-6
Non-life-supporting equipment
3 V
150 kHz to 80 Mhz
Life-supporting equipment
3 V outside ISM band
10 V inside ISM band
Electrical fast transients/burst
2 kV for power supply lines
1 kV for input/output lines > 3 m
1 kV differential mode, 2 kV
common mode
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
EN 6 1000-4-11
0% U
n
for 0.5 cycle
40% U
n
for 5 cycles
70% U
n
for 25 cycles
0% U
n
for 5 sec
Power frequency (50/60 Hz)
magnetic field EN 61000-4-8