PRODUCT FEATURES
32bit microprocessor, color display and graphing
Measures CO (ppm) & O2 (%)
Calculates COAF (air free) & EA (excess air)
Comes with a Water Trap, Particle & NOx filters
For residential combustion AND / OR ambient applications
Strong magnet for easy mount on appliances
Rechargeable Lithium-ion battery via easy USB charger
Professional WATER PROOF hard-case
Gas Sample Pistol Probe
PRODUCT COMPONENTS
1. Analyzer - CO & O2 sensors with a niobium magnet for mounting.
2. Sample Gun with Stainless Steel Probe.
3. Hose assembly. Silicone tubing, Water trap, particle and NOx filters. The NOx filter
removes acid gases to minimize CO sensor interference.
4. Waterproof Hard-case
5. USB Charging cable
6. USB Charger
7. Calibration NIST Certificate
8. Manual (this document)
IMPORTANT OPERATIONAL NOTES
- Ensure the analyzer is stored well within specifications.
- When using the analyzer, condensation may occur and the water trap will
slowly fill with water. Keep track of this. Empty as necessary. Always empty
before storing and retiring the unit.
- The sensors have a rated life of 2 years. If well taken care of, they can last
longer (up to 3 years) but will require more frequent calibration.
- Please email FORENSICS DETECTORS if any part requires replacement.
- When placing the probe in the exhaust flue, ensure it remains in the hot
exhaust for no longer than 2 minutes. Avoid high temperatures for
prolong period of time (<3 minutes). The pistol gun made get VERY HOT.
- Gas sample probe may get hot. BEWARE of burning your hand.
- If unsure, always ask. Email us: sarah@forensicsdetectors.com