- 22 -
APPENDIX III
Recommended separation distances between
portable and mobile RF communications equipment and the Sphygmomanometer
(
MD2200
)
The Sphygmomanometer (
MD2200
) is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Sphygmomanometer (
MD2200
) can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
Sphygmomanometer (
MD2200
) as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output power of
transmitter (W)
Separation distance according to frequency of transmitter (m)
10 V/m
0.01
0.1
1
10
100
0.200
0.632
2.000
6.33
20.0
0.060
0.190
0.600
1.90
6.00
3 Vrms
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.