5.2 Support
LED indications
Situation
Description
Connected via
USB
Green continuous
Orange blinking
Orange blinking
Red continuous
During measure
Green blinking (2 flashes per blink)
Green blinking (2 flashes per blink)
Green/Orange blinking (1 green, 1 orange flash per
blink)
Error
2 rapid red blinks
3 rapid red blinks
4 rapid red blinks
5 rapid red blinks
6 rapid red blinks
Error type
Error type*
How to solve the issue
Error 1
Disconnect and replug the Physilog
®
5
Error 2
Contact GaitUp
Error 3
Charge the Physilog
®
5
Error4
Contact GaitUp
Error 5
Contact GaitUp
Error 6
Empty the memory card
*see LED indication above
Soft reset of Physilog
®
5
The soft reset is the first manipulation which can be done if the Physilog
®
5 is not working
properly. To do a soft reset, hold the main button for 15 seconds. The LED shines orange
and then stops. If the problem persists, fill out the contact form on the support page
(www.gaitup.com/support) or write an email to contact@gaitup.com. Please indicate
Physilog
®
5 serial number and describe the problem in as much detail as possible.