10 K0554 Revision A
Figure 8 Product labels
5 Operating Environment
WARNING
THIS EQUIPMENT IS NOT RATED FOR USE IN POTENTIALLY EXPLOSIVE
ATMOSPHERES.
Storage temperatures ADTS542F: -20°C to 70°C (-4° to 158°F)
ADTS552F )
ADTS553F ) -30°C to 70°C (-22° to 158°F)
ADTS554F )
Operating temperatures/
environment
ADTS542F: 0°C to 50°C (32° to 122°F)
ADTS552F )
ADTS553F ) 0°C to 50°C (32° to 122°F)
ADTS554F )
Operating humidity: 5% to 95% RH (condensing)
Indoor & Outdoor use
ADTS Touch universal
power supply
Operating temperature: 10°C to 50°C (50° to 122°F)
Operating humidity: 5% to 95% RH (non-condensing)
Power supply Universal input Mains Power Supply.
Class 1 equipment requiring a protective earth connection.
TXAH 250V HRC Time-delay (T) 250Va.c. rated fuses (X = amp
rating of fuse).
Overvoltage category II, Pollution degree 2
The supplied mains/PSU cable protective earth/ground con-
ductor must be connected to the power supply protective
earth/ground system.
Electrical safety EN 61010-1, UL61010-1, CSA 22.2, No. 61010-1 and
IEC61010-1
Operating altitude Maximum 7500ft (2300 metres)
Ingress protection ADTS542/552/553/554 IP23, ADTS TOUCH IP65 (EN60529)