23
Rear Pedestal is newly designed.
1. The 16ch Switch is not used in SV system, which is functionally replaced by Mega Switch
located on the Magnet Side kit.
2. LED Power Box is introduced, which is the power supply for SRI, PAC, Patient alignment
laser and bore light.
3. Dummy load is moved to rear pedestal from the LPCA.
4. Hybird, TNS antenna, motor, encoder assembly, encoder box and communication box
have no change. They are same as the parts in HD system.