Manufacturer | GE Healthcare |
---|---|
Model | Carestation 650c |
Power Supply | 100-240V AC, 50/60 Hz |
Oxygen Concentration | 21-100% |
Category | Medical Equipment |
Display | 15-inch color touchscreen |
Battery Backup | Approximately 90 minutes (depending on configuration and usage) |
Connectivity | Ethernet, USB |
Ventilation Modes | Volume Control, Pressure Control, SIMV |
Tidal Volume Range | 20-1500 mL |
Pressure Range | 5-60 cmH2O |
PEEP | 0-20 cmH2O |
Integrated Monitoring | Yes, with respiratory and gas monitoring |
Oxygen Flush | Yes |
Vaporizers | Supports up to 3 vaporizers (optional) |
Ensure the pendant lift can support the anesthesia machine's total weight, recording the verification.
Verify minimum distance from lift pins to ground is less than 550mm (21.65 inches).
Remove packaging and any extra loads from the anesthesia machine's top shelf and tabletop.
Loosen tilt blocks, slide them, and turn the tilt adjustment screw counterclockwise.
Align lift pins with machine mount holes and raise lift until fully mated.
Position tilt blocks against the bracket, tighten screws to torque, and reinstall parts.
Connect the power cable and gas supplies to the anesthesia machine.
Execute the anesthesia machine Checkout Procedure as detailed in the Technical Reference Manual.
Lower the pendant lift so the machine rests on wheels and lift pins retract from mounting holes.
Move the anesthesia machine away from the pendant lift and perform service activities.
Verify tightness of sixteen screws during maintenance, checking alignment of adjacent areas.
Remount the anesthesia machine to the pendant system following the installation instructions.