GE HEALTHCARE LOGIQ C3/C5 PREMIUM
DIRECTION 5341787-100, REVISION 8 BASIC SERVICE MANUAL
Chapter 4 Functional Checks 4 - 3
4-3-2-1Back-end Processor Power Down (cont’d)
2.) The System-Exit window is displayed.
Figure 4-20 System Exit Window
3.) Using the Trackball or Select key, select Shutdown.
4.) The shutdown process takes a few seconds and the power off sequence is complete when the
power status LED is turned amber.
5.) Disconnect the probes.Clean or disinfect all probes as necessary. Store them in their shipping
cases to avoid damage.
4-3-2-2 Scanner Shutdown
Disconnect the Mains Power Cable is necessary. For example: Relocating the scanner.
CAUTION
DO NOT unplug and/or transport the unit until after the power off sequence has been completed.
Failure to do so may result in corrupted patient files.