EasyManuals Logo

GE MAC 1600 Manual

GE MAC 1600
8 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #7 background imageLoading...
Page #7 background image
stay productive.
Application options
Stress testing application (ergometer and treadmill
sold separately)
Ergometers supported include: eBike, Ergoline 900,
Variobike, Excalibur
Treadmills supported include: T2100, T2000, Trackmaster
TMX425
Specifi cations
Height / width / depth: Approximately
400 mm x 350 mm x 150 mm
Weight: Approximately 6.0 kg with battery installed
Power supply
• AC operation
Voltage: 100 to 240 VAC, +10, -15%
Current: 0.5A at 115 VAC
Current: 0.3A at 240 VAC
Frequency: 50 to 60 Hz +/- 10%
• Battery operation
Type: User replaceable, 16.4V@ 2300 AH, Lithium Ion
Capacity: 50 single page reports, (typical) 3 hours
continuous display (without printing)
Charge time: Approximately 2.5 hours from total discharge
Temperature
Operating: 50° to 104° F (10° to 40° C)
Transport/storage: -40° to 158° F (-40° to 70° C)
Humidity
Operating: 20% to 95% RH non-condensing
Transport/storage: 15% to 95% RH non-condensing
Pressure
Operating: 700 to 1060 hPa
Transport/storage: 500 to 1060 hPa
Warranty:
3 years for ECG device, 90 days for patient cable and battery.
Certifi cations:
EN 60601-1: 1990, A1: 1993, A2: 1995, A13: 1996 (IEC
60601-1: 1998, +A1: 1991, +A2: 1995) Medical electrical
equipment – Part 1: General Requirements for Safety
IEC 60601-1-1: 2000 Medical Electrical Equipment:
General Requirements for Safety
IEC 60601-1-2: 2007 General Requirements for Safety
Electromagnetic Compatibility
IEC 60601-1-4:2000 Requirements for Collateral Standard
Programmable Electrical Medical Systems
IEC 60601-2-25: 1993 +A1: 1999 Safety of Electrocardiographs
IEC 60601-2-51: 2003 Safety and performance
of ECG recorders
ANSI/AAMI EC11: 1991 Diagnostic Electrocardiograph Devices
JIS T 0601-2-25:2006 Japanese Industrial Standard
for Electrocardiographs
UL 60601-1: 2006 UL Standard for Safety Medical Electrical
Equipment, Part 1: General Requirements for Safety
CAN/CSA C22.2 No. 601.1
ANSI/AAMI EC57: 1998: Testing and reporting performance
results of cardiac rhythm and ST segment measurement
algorithms.

Other manuals for GE MAC 1600

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC 1600 and is the answer not in the manual?

GE MAC 1600 Specifications

General IconGeneral
BrandGE
ModelMAC 1600
CategoryMedical Equipment
LanguageEnglish

Related product manuals