Care and Maintenance
4-40 Ultrasound System – Common Service Information
Direction 5444964-100 English
 Rev. 5
Isolated Probe (Sink) Leakage-Isolation Test
Measurements should be recorded for probes and transducers 
under each set of test conditions specified in: 
• Table 4-8 on page 4-21
Record all data and keep the record of the results with other 
hard copies of maintenance data. 
1.  Connect the Safety analyzer to an AC wall outlet. 
2.  Plug the equipment under test into the receptacle on the 
panel of the Electrical Safety Analyzer. 
3.  Connect the Ultrasound probe to be tested to the Ultrasound 
system. 
4.  Immerse the saline probe in the saline solution. 
5.  Immerse the Ultrasound probe’s face (imaging area of the 
probe) into the saline solution. 
6.  Select the “Patient Lead Leakage” function on the Electrical 
Safety Analyzer. 
7.  Test with closed ground with the Ultrasound system on and 
off. 
NOTE:  For more information, refer to the safety analyzer's user manual.
WARNING
Do not use the probe if the insulating material has been 
punctured or otherwise compromised. Integrity of the insulation 
material and patient safety can be verified by safety testing 
according to IEC60601-1.
CAUTION
Line voltage is applied to the ultrasound probe during this test. 
To avoid possible electric shock hazard, the system being 
tested must not be touched by patients, users or anyone while 
the ISO TEST switch is depressed.