Facility needs
Ultrasound System – Common Service Information 3-25
Direction 5444964-100 English Rev. 5
Required facility needs  (continued)
•  Dedicated single branch power outlet of adequate 
amperage, meeting all local and national codes, which is 
located less than 2.5 m (8 ft.) from the Ultrasound system’s 
proposed location
•  Door opening is at least 76 cm (30 in) wide
•  Proposed location for Ultrasound system is at least 0.5 m 
(1.5 ft.) from the wall for cooling
•  Power outlet and place for any external peripheral are within 
2 m (6.5 ft.) of each other with peripheral within 1 m of the 
Ultrasound system to connect cables.
•  Power outlets for other medical equipment 
•  Power outlets for test equipment within 1 m (3.2 ft.) of 
Ultrasound system
•  Clean and protected space to store probes (in their cases or 
on a rack)
•  Material to safely clean probes (done with a plastic 
container, never metal)
For the amperage requirements, see: ‘Electrical requirements’ 
on page 3-5.
Desirable features
•  Door is at least 92 cm (3 ft.) wide
•  Circuit breaker for dedicated power outlet is easily 
accessible
•  Sink with hot and cold water
•  Receptacle for bio–hazardous waste, like used probe 
sheaths
• Emergency oxygen supply
•  Storage for linens and equipment
•  Nearby waiting room, lavatory, and dressing room
•  Dual level lighting (bright and dim)
•  Lockable cabinet ordered by GE for its software and 
proprietary manuals