Electrical safety tests
Ultrasound System – Common Service Information 4-25
Direction 5444964-100 English Rev. 5
Chassis leakage current test
Definition
This test, also known as Enclosure Leakage current test, 
measures the current that would flow through a grounded 
person who touches the accessible conductive parts of the 
equipment during normal and fault conditions. 
The test verifies the isolation of the power line from the chassis.
The testing Electrical Safety Analyzer is connected to parts of 
the equipment, easily contacted by the user or patient. 
Record the highest reading.
DANGER Electric Shock Hazard. 
When the Electrical Safety Analyzer's ground switch is OPEN, DO NOT 
touch the Ultrasound system!
CAUTION
Equipment damage possibility. 
Never switch the Polarity and the status of Neutral when the 
Ultrasound system is powered ON.  
Power off the Ultrasound system, allow the stored energy to 
bleed down, and turn the circuit breaker off BEFORE switching 
the "POLARITY" switch and/or the "NEUTRAL" switch on the 
Electrical Safety Analyzer to avoid possible power supply 
damage.