GE MEDICAL SYSTEMS - KRETZTECHNIK ULTRASOUND
D
IRECTION 105844, REVISION 1VOLUSON® 730 SERVICE MANUAL
Chapter 3 Installation 3-1
Chapter 3
Installation
Section 3-1
Overview
3-1-1 The Purpose of Chapter 3
This chapter contains information needed to install the unit. Included are procedures to receive, unpack
and configure the equipment.
3-1-2 Average Installation Time
Manpower requirement: 2 persons
Table 3-1 Contents in Chapter 3
Section Description Page Number
3-1
Overview
3-1
3-2
Receiving and Unpacking the Equipment
3-4
3-3
Preparing for Installation
3-6
3-4
Completing the Installation
3-8
3-5
Printer Installation
3-12
3-6
System Configuration
3-28
3-7
Connectivity Installation Worksheet
3-36
3-8
Network IP Address Configuration
3-37
3-9
Paperwork
3-38
Table 3-2 Average Installation Time
Description Average Installation Time Comments
Unpacking the scanner 0.5 hours
Scanner /options / printers 0.5 to 1.5 hours Dependant on the required configuration
DICOM Option 0.5 - 1.5 hours Dependant on the configuration amount