GE MEDICAL SYSTEMS - KRETZTECHNIK ULTRASOUND
D
IRECTION 105844, REVISION 1VOLUSON® 730 SERVICE MANUAL
Chapter 3 Installation 3-3
3-1-4 Safety Reminders
DANGER
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH
THE UNIT!
CAUTION
Two people should unpack the unit because of its weight. Two people are required whenever
a part weighing 16kg (35 lb.) or more must be lifted.
CAUTION
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate
to its operating environment.
CAUTION
To prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use
a three to two prong adapter. This defeats safety grounding.
CAUTION
Do NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is
present.
CAUTION
Do not use a20 Amp to 15 Amp adapter on the 120 Vac unit’s power cord. This unit requires
a dedicated 16 A circuit.
CAUTION
Do not operate this unit unless all board covers and frame panels are securely in place.
System performance and cooling require this.
CAUTION
OPERATOR MANUAL(S)
The User Manual(s) should be fully read and understood before operating the Voluson® 730
and kept near the unit for quick reference.
CAUTION
ACOUSTIC OUTPUT HAZARD
Although the ultrasound energy transmitted from the Voluson® 730 probe is within FDA
limits, avoid unnecessary exposure. Ultrasound energy can produce heat and mechanical
damage.
Figure 3-2 Environmental Labels