GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 4 - Functional Checks 4-15
4-4-6 Volume Mode Checks
NOTE: Different menus are displayed depending on which Touch Panel menu and which Volume Mode
is selected.
NOTE: Some function keys only appear on the Touch Panel if they are available for the selected Probe.
4-4-6-1 Pre-Volume Mode Functions
NOTICE
!! NOTICE:
Real Time 4D, RT_4D_Biopsy, DiagnoSTIC, VCI and VOCAL are Options.
If these options are not part of the system configuration, the appendant checks can be omitted.
Figure 4-12 Pre-Volume Mode menus (e.g., Static 3D Render)
1. Touch this key to select desired Acquisition- and
Visualization Mode.
2. Select Acquisition Mode
3. Select Visualization Mode
4. Select User Setting
5. Start the Volume Acquisition with the Freeze key resp. the right trackball key.