GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 4 - Functional Checks 4-31
4-5-2 Load User Settings Only (Application Settings)
1.) Insert the DVD/CD+(R)W or MO (Magneto-Optical Disk) into the drive.
2.) On the Touch Panel, touch UTILITIES
.
3.) In the Utilities menu, touch SYSTEM SETUP
to invoke the setup desktop on the screen.
4.) Select the
BACKUP page and click the LOAD button of the “User Settings Only” group.
5.) Choose the media (e.g., DVD/CD+RW) and click the LOAD
key.
6.) Select the appropriate file and click OK
. The “Load Backup Data” window appears.
CAUTION
!! CAUTION:
The loading procedure overwrites the existing application settings on the local hard drive. Make sure to
insert the correct DVD/CD or MO. Additionally you can load the backup from D:\UserSettings.
Figure 4-24 User Settings Only - Load window
NOTICE
!! NOTICE:
It is highly recommend to use Application settings which are adapted for the systems software version!
Figure 4-25 Load Backup Data
Choose the media
and
then click [Load]
Complete Backup
[>>] button
[Load] button