4-5-3 Save Full Backup (Presets, Configurations & Application Settings)
A full backup always contains the following data
Patient demographic and exam data (database containing the patient data and measurements)
SonoView image data (NOT available when saving to the internal hard disk, DVD/CD or MOD)
User Settings (databases and files containing gray curves and the user settings.)
Image transfer settings (DICOM settings e.g., DICOM servers, AE Title, Station Name, etc.)
Measure Setup Settings (user specific measure settings)
V730 settings (general settings such as language, time/date format and the enabled options)
Windows Network Settings (network settings including the computer name)
Service Platform (state of the service platform)
VP (additional system data)
NOTE: Always “Full Backup” any presets, configurations and application settings to HDD and/or DVD or MO-
disk before upgrading the software and/or application settings. This ensures that if the presets need to
be reloaded, will be the same ones the customer was using prior to service.
1.) On the Touch Panel, touch UTILITIES
.
2.) In the Utilities menu, touch SYSTEM SETUP
to invoke the setup desktop on the screen.
3.) Select the
BACKUP page and click the SAVE button of the “Full Backup” group.
4.) Choose the Destination (e.g., Network Drive).
5.) Enter the description of the full backup.
6.) If desired and possible (Network Drive and Other drive only), activate “Include Images.
NOTE: The “Include Images” option may result in a large amount of data: up to 70 Gigabytes!
7.) Select the NEXT
button to start the backup process.
After copying the data, the Voluson® 730Expert reboots and the application starts again.
CAUTION
!! CAUTION:
It is recommended to create a full backup of the settings once a week.
Figure 4-27 Full Backup Save
choose destination
enter description
click [Next] button