GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
5-24 Section 5-4 - FrontEnd Processor
5-4-1-4 CRW - CW-Doppler Board (optional)
Separate Clock Oscillator + Clock generator PRF, FN
•TxDelayMatrix
Transmitter,
Receiver Amplifiers (RxAmp),
Receiver Delay matrix (RxDelayMatrix)
Analogue: Mixer, Amplifier, Filter (WMF, Anti Aliasing)
PRF-SampADC16Bit,
5-4-1-5 CPZ - Cover Board
Transfers the analogue transmitting / receiving signals between the following boards:
•CPU
•CPR
CRW
192 Transmitter-, 128 Receiver channels, 16 CW-Doppler channels - switched by relays at CPZ
5-4-1-6 CPK - Motherboard of GEF-Module
Following boards are direct connected to the CPK:
CPU - Probe connector Board
CPR - Beam former Board
CRW - CW-Doppler Board
CRS - Signal Processing Board
CPP - Power Supply Board
CPM - Motherboard (electrical Signal- and Supply connection between all PC-Plug-In Boards)