GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
5-30 Section 5-5 - BackEnd Processor
5-5-2-7 CPP - Power Supply Secondary Board + Motor Power stage
used for Supply of both FrontEnd and BackEnd
DC-DC-Converter:
59VDC to following output voltages:
+ 3.3V, +/-5V, +12V, +/-15V,
+ FAN (10 -24V / 15 Watt; adjustable by software)
+TX_POW (+/-90V)
-TX_POW (+/-90V)
Motor Sinus2 Powerstage Amplifier
Motor Cosinus2 Powerstage Amplifier
•+TX_POW2
•-TX_POW2
see also: Section 5-4-1-8 on page 5-26