GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
5-40 Section 5-10 - Peripherals
Section 5-10
Peripherals
5-10-1 General Information - GEM (Disk Drive Module)
The GEM Module contains the DVD/CD+(R)W Drive.
Additionally the Magneto Optical Drive and the ECG-preamplifier (MAN6) can be installed as an Option.
5-10-1-1 ECG-preamplifier (MAN6 - optional)
The ECG-preamplifier is used for acquiring an ECG-signal to be displayed with the ultrasound image.
This optional peripheral serves for gaining an ECG-signal to mark the systolic and end diastolic
moments in M-Mode and Doppler evaluations.
The ECG-preamplifier must not be used for ECG-diagnostics. It is not intended for use as a cardiac
monitor and must not be used for an intraoperative application on the heart.
5-10-1-2 DVD/CD+(R)W Drive
The DVD/CD+(R)W Drive (Writer) is used to backup images and reports. In addition, it is used as the
main source of software upgrades and other service utility operations. The DVD/CD+(R)W Drive can
read/write CD´s and DVD´s.
5-10-1-3 Magneto-Optical Drive (optional)
Storage capacity by disk: 1.3GB, 640MB, 540MB, 230MB, 128MB
The MO-Drive allows to read and write any GIGAMO standard 1.3GB disk at twice the liner bit density.
Additionally it retains full read/write compatibility with ISO/IEC 3.5 - inch disks ranging from original
128MB to current 640MB.
MO disks are nearly indestructible and immune from the problems that plague magnetic media.
MO disks can be rewritten an unlimited number of times.
Figure 5-23 GEM incl. optional MOD and ECG
optional ECG (MAN6)
DVD+(R)W Drive
optional
Magneto-Optical Drive (MOD)