GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
5-52 Section 5-15 - Service Page
4.) Enter the password SHE and click the ACCEPT button to display the Service Tools window.
5-15-3-1 Auto Tester
Autotest is a log function of customer activities. It records all user actions (scanning, Touch Panel
entries, performing Calculations, review of Patient Reports, etc.…). It is possible to safe (record) as file
on HDD. But also export to DVD/CD or MOD can be done to allow replay of the records on other units.
Operation see: Section 7-5 "How to use the Auto Tester program" on page 7-7.
Figure 5-37 Service Tools window
NOTICE
!! NOTICE:
For intermittent problems this file can be requested from the Service Engineer or customer.
It is possible to burn this file on DVD/CD+(R)W or to store it on MO-disk.
Service Viewer
Maintenance Report
Export System Data
Auto
Tes ter
EUM
Monitor Test
Delete all patients
Printer
UIS
SSW