GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 6 - Service Adjustments 6-7
Section 6-6
Daylight Saving Time (DST) - New Dates
Procedure:
1.) On the Touch Panel, touch UTILITIES.
2.) In the Utilities menu, touch SYSTEM SETUP
to invoke the setup desktop on the screen.
3.) Select the
GENERAL page in the System Setup.
4.) Click the DATE/TIME
button.
5.) In the displayed “Date/Time Properties” window select the
TIME ZONE tab.
6.) Uncheck “Automatically adjust clock for daylight saving changes” check box and then click APPLY
(see: left image of Figure 6-6 below).
7.) Select the
DATE / TIME tab in the “Date/Time Properties” window.
8.) Set the proper time for the system’s location and then click OK
(right image of Figure 6-6 above).
9.) Click the SAVE&EXIT
button to save your changes and exit the System Setup page.
The System is shutting down automatically and restarts again.
Issue: The automatic Daylight Saving Time feature in Microsoft Windows products may no longer run on the
correct date, causing the system’s time to be incorrect.
Cause: Some countries have changed their start and end dates for Daylight Saving Time (DST).
Access Microsoft website http://support.microsoft.com/kb/928388 to determine if the system’s location
site is affected.
Solution: If the Voluson® 730Expert relies on the automatic DST feature in Microsoft Windows, turn off the
feature and then set the correct system time manually at the start of DST and again at the conclusion
of DST.
Figure 6-6 Uncheck check box / adjust time