GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
7-8 Section 7-5 - How to use the Auto Tester program
Section 7-5 How to use the Auto Tester program (cont’d)
The following message box appears.
5.) Click OK
.
6.) Press the PAUSE
key on the keyboard.
The Auto Tester window appears, see: Figure 7-7 below.
7.) Activate the “Auto Tester” program by clicking the “Record” icon on the displayed screen.
8.) Start scanning.
You can scan normally and everything will be recorded to the program (up to several hours.)
NOTE: It is important that you are recording the processes where the errors normally occur.
9.) Stop the program by pressing the PAUSE
key on the right upper corner of the keyboard.
Figure 7-6 Message Box
Figure 7-7 Start Auto Tester
NOTICE
Please check status of the “Accumulate mouse moves” function. It has to be turned OFF.
i.e., no check mark visible!
Pause
key on the
right upper corner
of the keyboard
“Record”
icon
“Accumulate mouse
moves” function has
to be turned OFF
(no check mark)