EasyManua.ls Logo

GE Voluson 730Expert - Section 7-7 Troubleshooting Trees and Instructions

GE Voluson 730Expert
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-11
Section 7-7
Troubleshooting Trees and Instructions
7-7-1 System does not Power On / Boot Up
Figure 7-12 System does not Power On / Boot Up
Is the scanner plugged
in?
Is there AC at the wall
outlet?
Make sure the
Scanner isn't
plugged into
emergency power
(
red outlet
)
Yes
Is the circuit breaker on
Yes
+ These three condidtions should be the first items checked if
the scanner doesn't boot up.
+ If all three conditions are present, the scanner should be in
a standby mode with the ON/OFF switch illuminated with amber
color.
+ Follow the flow chart to determine the reason the scanner
doesn't boot up.
No
No
Switch in ON
Position
Fan Noise ?
Yes
+ Check if the power switch over the mains plug of the Primary Power
Supply isn't switched off.
+ Check the connections to and from the Power Supply.
+ Check the power of the cable which connects the Power Supply with
the GEF module. (description Check Points Voltages)
+ If ther is no power and the connections are O.K., replace the Primary
Power Supply.
No
Yes
System is on and ready for use
Section 7-3