EasyManua.ls Logo

GE Voluson 730Expert - Page 226

GE Voluson 730Expert
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
7-16 Section 7-7 - Troubleshooting Trees and Instructions
7-7-6 Unable to Record to VCR
Figure 7-16 VCR - Troubleshooting
Unable to record to VCR!
Check that all signal, and power
supply cable connections to the VCR
unit are OK.
YES
NO
Is the VCR properly
connected?
Is the VCR tape put into
the device?
Put the tape into the device and
rewind it.
NO
YES
Replace VCR
Is recording unable even
by manual operation?
YES
- Check the signal cable connections
between VCR and Internal I/O.
NO
YES
Video Output of System
o.k. ?
Replace CKV
NO

Related product manuals