GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-25
File Pos load volume files from other storage medium
File Read Error load volume files from other storage medium
File Type Unknown load volume files from other storage medium
File Volume size not consistent load volume files from other storage medium
FLT: Denormal operand restart the system
FLT: Divide by zero restart the system
FLT: Invalid operation restart the system
FLT: Stack overflow restart the system
FLT: Underflow restart the system
FrameUp - TempBuffer: Memory allocation failed restart the system
GeoDescription3D_TissueCF::operator=: handed over
argument is not of same type, dynamic cast failed!!
restart the system
GeoDescriptionMotion Constructor: unknown mode Parameter
handed over
restart the system
GeoDescriptionMotion::operator=: handed over argument is not
of same type, dynamic cast failed!!
restart the system
GeoDescriptionPWMode::operator=: handed over argument is
not of same type, dynamic cast failed!!
restart the system
GeoDescriptionTissueCF Constructor: unknown mode
Parameter handed over
restart the system
GeoDescriptionTissueCF::operator=: handed over argument is
not of same type, dynamic cast failed!!
restart the system
Hardware doesn't support CW-mode pencil probe + CW-Hardware not available->HW problem
hardware error on restart the system
HardwareRelatedSoftware_Windows in write have different
ProbeAcousticUnitIDs
disconnect all connected probes and connect them again;
if not ok restart the system
HardwareRelatedSoftware_Windows in write have different
ProbeScanFuncIDs
disconnect all connected probes and connect them again;
if not ok restart the system
IBegrenzer.cpp Bshots TxMultiFocus problem restart the system
IBegrenzer.cpp Mshots TxMultiFocus problem restart the system
In the 3D Image Measure is not allowed change to another format than 3D Fullscreen mode
incorrect VersionByte, xx of xx cycles OK contact technical support
iSetVideoSource(eVideoIntern) function fails reboot the system
iSetVideoSource(eVideoIntern) function fails the system will restart itself by pressing OK
LineFilt Out Of Range press ok and save this user-setting once again
LP_KoefBlock: SamplePRF too big restart the system
Error Messages Actions