8-8 Section 8-3 - User Settings Only (Application Settings) Loading Procedure
Section 8-3
User Settings Only (Application Settings) Loading Procedure
8-3-1 Introduction
The User Settings contains:
• Application Settings
•User Programs
•Auto Text
• 3D/4D Programs
8-3-2 Loading Procedure
see: Section 4-5-2 "Load User Settings Only (Application Settings)" on page 4-31
Section 8-4
Full Backup (Presets, Configurations & Appl. Settings) Loading Procedure
8-4-1 Introduction
The Full Backup contains following data:
• Patient demographic and exam data (database containing the patient data and measurements)
• SonoView image data (NOT available when saving to the internal hard disk, DVD/CD or MOD)
• User Settings (databases and files containing gray curves and the user settings.)
• Image transfer settings (DICOM settings e.g., DICOM servers, AE Title, Station Name, etc.)
• Measure Setup Settings (user specific measure settings)
• V730 settings (general settings such as language, time/date format and the enabled options)
• Windows Network Settings (network settings including the computer name)
• Service Platform (state of the service platform)
• VP (additional system data)
8-4-2 Loading Procedure
see: Section 4-5-4 "Load Full Backup (Presets, Configurations & Application Settings)" on page 4-35
NOTICE
!! NOTICE:
From Software Version 3.1.0 onwards, it is possible to load a previously saved “Full Backup” into the
Voluson® 730Expert.