GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
10-4 Section 10-3 - Maintenance Task Schedule
* Twice a year is recommended.
Console Leakage
Current Checks
Also after corrective
maintenance or as required by
your facilities QA program.
Peripheral Leakage
Current Checks
Also after corrective
maintenance or as required by
your facilities QA program.
Surface Probe Leakage
Current Checks
Also after corrective
maintenance or as required by
your facilities QA program.
Endocavity Probe
Leakage
Current Checks
•*
Also after corrective
maintenance or as required by
your facilities QA program.
Measurement Accuracy
Checks
Also after corrective
maintenance or as required by
your facilities QA program.
Probe/Phantom Checks
Check axial and lateral resolution (see Basic
User Manual Technical specifications).
Check Gain and TGC changes, vary the
focus and check reaction on screen.
Also after corrective
maintenance or as required by
your facilities QA program.
Functional Checks
of all probes Section 10-
5-2 on page 10-7
Also after corrective
maintenance or as required by
your facilities QA program.
Table 10-2 Customer Care Schedule
Item Service at Indicated Time D W M A Notes