GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 10 - Care & Maintenance 10-23
10-7-9-4 Data Sheet for Transducer Source Leakage Current
The test passes when all readings measure less than values shown in Table 10-12 and Table 10-13.
Record all data on the Ultrasound Inspection Certificate (see: page 10-25).
.
CAUTION
!! CAUTION:
When the units power is ON, never switch the Polarity and the status of Neutral. Be sure to turn
OFF power to the unit before switching them, using the POLARITY switch and/or the NEUTRAL
switch. Failure to comply with this warning may cause damage to the unit!
Table 10-19 Typical Data Sheet For Transducer Source Leakage Current
Transducer Tested:
Unit Power
Tester Power Polarity
Switch
Tester GROUND or
NUETRAL Switch
Measurement
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED