GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 1 - Introduction 1-11
1-4-2 Compliance
The Voluson® 730Expert unit conforms to all applicable conducted and radiated emission limits and to
immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line
transient requirements.
NOTE: For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh, clamps, are in
good condition, installed tightly without skew or stress. Proper installation following all
comments noted in this service manual is required in order to achieve full EMC performance.
1-4-3 Electrostatic Discharge (ESD) Prevention
Section 1-5
Customer Assistance
1-5-1 Contact Information
If this equipment does not work as indicated in this service manual or in the Basic User Manual, or if
you require additional assistance, please contact the local distributor or appropriate support resource,
as listed below.
NOTE: Prepare vital system information (see: Section 7-2 on page 7-2) before you call:
System Type
System Serial number (also visible on label on back of the system)
Application Software version
Backup version
additional information about installed software
WARNINGWARNING
!! WARNING:
DO NOT touch any boards with integrated circuits prior to taking the necessary ESD
precautions:
1.) When installing boards, ESD may cause damage to a board.
ALWAYS connect yourself, via an arm-wrist strap, to the advised ESD connection
point located on the rear of the system (to the right of the power connector).
2.) Follow general guidelines for handling of electrostatic sensitive equipment.
WARNINGWARNING
!! WARNING:
Risk of electrical shock, system must be turned off.
Avoid all contact with electrical contacts, conductors and components.
Always use non-conductive handles designed for the removal and replacement of ESD
sensitive parts. All parts that have the potential for storing energy must be discharged
or isolated before making contact.