GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
Chapter 2 - Site Preparation 2-5
2-2-4 Probe Environmental Requirements
Operation: Ambient temperature 18° to 30° C
Storage: -10° to 50° C
NOTE: Temperature in degrees C. Conversion to degrees F = °C (9/5) + 32).
2-2-5 Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
2-2-6 System Specifications
2-2-6-1 Physical Dimensions of Voluson® 730Expert
The physical dimensions of the Voluson® 730Expert unit are summarized in Ta ble 2-5 .
Table 2-6 lists the size of Voluson® 730Expert, with monitor and without on-board peripherals.
2-2-6-2 Weight without Monitor and Peripherals
2-2-6-3 Acoustic Noise Output
max. 57dB(A)
2-2-6-4 Electrical Specifications
Please refer to Section 2-2-2-1 "Voluson® 730Expert Power Requirements" on page 2-3.
NOTICE
!! NOTICE:
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
-10 TO + 50 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
CAUTION
!! CAUTION:
Have two people available to deliver and unpack the Voluson® 730Expert.
Attempts to move the unit considerable distances or on an incline by one person could result
in injury or damage or both.
Table 2-5 Physical Dimensions of Voluson® 730Expert
Height Width Depth
142 cm / 55.9 inches 68 cm / 26.8 inches 100 cm / 39.4 inches
Table 2-6 Weight of Voluson® 730Expert with Monitor and without other Peripherals
Model Weight [kg] Weight [lbs.]
Voluson® 730Expert 136 300