GE HEALTHCARE - KRETZTECHNIK VOLUSON® 730EXPERT (BT03)
D
IRECTION 105899, REVISION 3 DRAFT (APRIL 29, 2008) SERVICE MANUAL
3-14 Section 3-5 - Connection of Auxiliary Devices
3-5-6-2 Sony SVO-9500MD
Figure 3-11 Sony - Video Recorder Connection Scheme
Figure 3-12 DIP Switches
NOTICE
!! NOTICE:
Please use the proper connection set and remote cable.
see: "Section 9-12 Optional Peripherals and Accessories (cont’d)" on page 9-28.
Set DIP Switches on back of the VCR as shown:
Switch 1: OFF-Audio is muted during fast playback
Switch 2: OFF-FF/REW mode is used during searches.
Switch 3: ON-VCR counter is reset upon cassette ejection.
Switch 4: OFF-The VCR can be controlled by the system.
Switch 5 + 6: ON-The Baud rate is set to 19200 bit/sec.
(must be the same as Voluson® 730Expert)