Site Preparations
3-24 Ultrasound System – Common Service Information
Direction 5444964-100 English
 Rev. 5
Required facility needs  (continued)
 Figure 3-2. Recommended floor plan, 4.27 x 5.18 m (14 x 17 foot)
Scale: Each square equal one square foot (app. 31 x 31 cm)
1.  Secretaries or Doctors Desk
2. File Cabinet
3. Film Viewer 
4. Counter Top
5.  Counter Top and Sink with 
hot and cold water
6.  Overhead Lights Dimmer - 
Dual Level Lighting (bright 
and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet - 
Circuit Breaker protected and 
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance 
of Ultrasound system from 
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and 
Equipment
16. Examination Table – 1930 x 
610 mm (76 x 24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 
inches)
4
1
2
18
5
6
9
16
17
15
8
7
10
11
12
3
13 14